- Visión General
- Product Description
- One-step Purchasing Solution
- Packaging & Shipping
- Company Profile
- After sales and Exhibitation
- Customer Feedback
- Contact Delia
Información Básica.
Descripción de Producto
Real Time Fluorescent Quantitative PCR Detection System




Product characteristics:
The risk of medical staff is small, reducing the probability of droplet infection.
It's fast and takes only 10 minutes to produces results.
No special testing environment and personnel training is required.
Compared with nucleic acid test swabs and pharyngeal swabs, it is less affected by samples and has high accuracy.
Dosage is less, only 1 drop of serum or plasma sample is needed.
Testing IgM and IgG at the same time to reduce the testing time.
Convenient, fast, suitable for mass screening.
Introduction:
We have developed powerful assay design algorithms, optimized qPCR regent, intuitive data analysis software, and flexible instrumentation to help harness the power of qPCR across a rich and diverse set of applications.
Explore our robust solutions for your qPCR-based research.
It can be widely used for Infectious disease research, Food pathogen detection, Waterborne pathogen detection, Pharmaceutical
analytics, Stem cell research, Pharmacogenomics research, Oncology and genetic disease research, Plant sciences and agricultural
biotechnology.
The temperature step change is controlled by the semiconductor peltier to realize PCR amplification.
Use high-sensitivity MPPC unit to detect fluorescence; Program control channel switching, non-contact excitation /detection on the
top structure, coordinated with motor control x and Y axis movement to achieve 96-hole scanning. Finally, accurate analysis is
carried out through powerful software.
When the human body is infected with the , the first antibody produced is the specific Igm antibody, which is generally produced after 3-7days of infection, and is replaced by a large number of specific IgG antibodies produced about two weeks later. Therefore, IgM and IgG of the specific antibodies can be used as an auxiliary means for the diagnosis of suspected patients.(Not Available in the U.S)
Product Name | Fluorescence Quantitative PCR Detection System | |||
Model | LEIA-X4 | |||
Sample Capacity | 96x0.1ml PCR plate, 12x8-strip tubes, 96*0.1ml single tube (Transparent Cover) | |||
Reaction System | 10~50 | |||
Dynamics Range | 1-1010 copies | |||
Channel | 4 | |||
Emission Light | LED | |||
Detector | MPPC | |||
Detection Path | F1 | F2 | F3 | F4 |
Suitable Probe/Dye | FAM/SYBR GREEN | VIC/JOE/ HEX/TET | ROX/TEXAS-RED | Cy5 |
Excitation Wavelength | 455~680nm | |||
Detection Wavelength | 510~730nm | |||
Fluorescence Detection Repeatability | CV≤2% | |||
Fluorescence Detection Accuracy | CV≤3% | |||
Fluorescence Detection Linearity | r≥0.995 | |||
Module Temp. Range | 4-99ºC(resolution:0.1ºC) | |||
Ramp Rate | 5.0ºC/s(max) | |||
Temp. Accuracy | ±0.3ºC | |||
Temp. Uniformity | ≤ ±0.3ºC | |||
Temp. Control Mode | Block mode | |||
Gradient Temp. Range | 1-36ºC | |||
Hot-Lid Temp. Range | 100ºC, Automatic Hot-lid | |||
Scanning Mode | Full plate scanning | |||
Programming | Max 100 Segments for Each Program, Max 99 Cycles | |||
Operation Mode | Continuous | |||
Scanning Time | 8.5s | |||
Special Function | Absolute quantitative automatic analysis, relative quantification, SNP Analysis, melting curve analysis, etc. | |||
Operation System | Microsoft: Windows10 | |||
Power Supply | 220V,50/60HZ; 110V,60HZ | |||
Dimension(L*W*H) Mm | 375*490 *365 | |||
Port Method | USB Port | |||
Packing Size(L*W*H) Mm | 645*565 *605 | |||
Gross Weight | 45 |